Home Agenda Archief Netwerk Nieuws | News Symposia Werkgroep

Program Abstracts Posters Submission Slides Hotels Directions
Downloads: |
FABIAN 2009
Socially
Responsible Bioanalysis
10 November 2009, Solvay
Pharmaceuticals, Weesp
|
A paradigm shift in
toxicological risk assessment: more science, less animal testing
Bas Blaauboer
(Universiteit Utrecht, NL)
The paradigm of toxicological risk assessment contains the elements of hazard
identification, hazard characterization, exposure assessment and risk
characterization. For different areas and applications of the process of risk
assessment, these elements have a different weight. The basis for the process of
determining risks is the ability to quantify toxicological endpoints, in most
cases in animal models. The next steps are the interpretation of the results,
assessment of the uncertainty in the data and extrapolation of the outcome of
the experimental findings to the risk situation one needs to assess, e.g. for
human health.
The heavy reliance on animal models in toxicity testing has prompted many
ethical issues. Furthermore, the costs of animal experimentation becomes a
serious burden. Also from a scientific point of view the usefulness of a risk
assessment relying on the apical endpoints in animal studies is doubted in many
cases.
Thus, the classical way of toxicological risk assessment needs redefinition in
the light of the modern paradigm of toxicology, viz. that toxicity is depending
on the proper dose metric (i.e. concentration or concentration in time) of the
proper compound (might be a metabolite) at the most critical site (i.e. the
target at which the primary mechanism occurs).
This implies that risk assessments will need to be more chemical-specific. This
will be possible if we can make use of data on the mechanism of toxicity in
combination with an estimation of the target concentration. The application of
new technologies such as the “omics”, a better use of in vitro methodologies and
of (physiologically-based) kinetic modelling will allow such an integrated
approach.
What is necessary is the integral use of data from the above-mentioned areas in
a transparent and logical way. Ideally, the new paradigm in toxicology would
then allow a safety testing for compounds without the use of animal
experimentation.
ALTERNATIVE TESTING STRATEGIES TO REDUCE ANIMAL USE IN PRECLINICAL
DEVELOPMENT AT J&JPRD BELGIUM
Bas-Jan Van der Leede (Johnson
& Johnson Pharmaceutical Research & Development, a division of Janssen Pharmaceutica N.V., Genetic and Exploratory Toxicology, Turnhoutseweg 30, Beerse,
BE)
Animal studies are still an essential part of the preclinical safety assessment
in the development of new pharmaceutical drugs. However the number of animals to
be used can be reduced significantly by applying different study designs and
introducing alternatives for animal testing. At J&JPRD several initiatives have
been implemented that contribute to animal welfare by replacement, reduction and
refinement of animal use in preclinical development of pharmaceutical drugs. An
additional advantage of the initiatives replacing animal testing is that
pharmaceutical drugs can be screened at higher throughput, thereby delivering
results faster. In order to identify the toxicity potential of new drugs
correctly, bio-analytical research can play an important factor in these
screening assays. The following initiatives will be presented in more detail.
Integration of in vivo micronucleus test (MNT) in rat
toxicity studies
The MNT and the 2-4 week rat toxicity study are both part of the preclinical
safely package submitted to regulatory agencies for authorization of
first-in-human trails. Integration of these studies saves approximately 120
animals per package.
In order to predict the potential human risk to develop cancer, the MNT needs to
be conducted at the highest possible systemic exposure of the drug. Since the
2-4 week rat toxicity study is typically conducted at lower doses, acceptance
criteria for dose selection are used at J&JPRD to decide whether integration of
the MNT is justified.
HET-CAM
Due to the fact that injection site reactions can result in serious problems
during preclinical i.v. infusion studies, there was a rising demand to develop
and implement an in vitro screening model to identify vascular toxicity prior to
animal testing. The HET-CAMVT(Vascular toxicity Hen’s Egg Test-Chorioallantoic
Membrane), developed by J&JPRD, is an in vitro model in which three irritation
effects (hemorrhage, coagulation and lysis) are evaluated after topical
application of a formulation on the CAM of fertilized chicken eggs. Experiments
in which formulations with known in vivo injection site reactions were applied,
showed that the HET-CAMVT model is a very sensitive assay, able to generate
reproducible and quantative results.
Zebrafish
Zebrafish have the potential to accelerate and de-risk the development process
by reducing attrition rates and lowering the development cost of new
pharmaceutical drugs. With a significant genetic similarity to humans and the
presence of many vital organs including heart, brain and liver, the larval
zebrafish is highly suitable for screening potential drug candidates for
efficacy and safety effects. At J&JPRD, zebrafish are used to predict
hepatotoxic properties of pharmaceutical drugs in development.
In vitro liver
toxicity predicts in vivo hepatoxicity using the microarray technique
Peter Olinga (Solvay
Pharmaceuticals, Clinical Candidate Selection, Weesp / University Groningen, NL)
We
have developed an in vitro system to study drug metabolism and toxicity in the
human and rat liver. Model compounds of liver toxicity were studied with rat
liver slices. The effect of these different hepatotoxins was analyzed at the
gene expression level in the rat liver in vitro. Microarray analysis in rat
liver slices correctly predicts in vivo hepatotoxicity. This in vitro technique,
the precision-cut liver slices, mimics the human and rat liver in vivo.
Optimal
kinetic screening (in-vivo and in-vitro) and evaluation
Jessica Dijksman (Solvay
Pharmaceuticals, Weesp, NL)
<< Abstract not available yet >>
Recent
Developments Towards Green Chromatography
Pat Sandra (Research Institute for
Chromatography (RIC), Kennedypark 26, B-8500 Kortrijk, B)
Chemists
are (should be) more and more concerned about the environmental impact of their
activities and at the end of the 20th century the term "green chemistry" was
introduced. The "green chemistry" principles should also be applied to
analytical chemistry and more especially to chromatography and its sample
preparation procedures.
Several analytical methods are, from an ecological point of view, very
unfriendly and, without any doubt, liquid chromatography (LC) is the first
technique to be considered. The present consumption of acetonitrile (an EPA
pollutant) in LC is enormous. Moreover, the waste disposal is incinerated
resulting in the formation of NO2 that is linked to acid rain.
The current trend of LC analyses is biased toward high throughput, high
productivity and high resolution. In response to these increasingly demanding
requirements, over recent years, innovative technologies and improvements in
instrumentation have emerged which are having a significant impact on our daily
work. Ultra High Pressure LC (UHPLC) and Elevated Temperature (ETLC) both have
extended speed, productivity and peak capacity for pharmaceutical,
environmental, food, and bio analyses. The application of both developments in
combination with a more or less forgotten technique i.e. sub- and supercritical
fluid chromatography, often allows to translate "unfriendly" methods into green
methods using as mobile phase constituents only water, ethanol and carbon
dioxide with formic acid and ammonia as additives. Several examples will be
shown to illustrate that conventionally used LC mobile phase compositions can
easily be replaced by these green alternatives.
The Use of Dried Blood
Spot Samples and Direct Analysis for the Quantitative Bioanalysis of Drugs
Paul Abu-Rabie (GlaxoSmithKline,
Bioanalytical Science and Development, UK)
The surge in interest in dried blood spot (DBS) techniques for supporting
pharmaceutical exposure studies is due to the many advantages it offers over
conventional plasma sampling. These include the reduction in blood volumes
required, with associated cost and ethical advantages, the simplification of
clinical sampling procedures and the reductions in sample processing, storage
and transportation costs. DBS analysis is now GlaxoSmithKline’s default
technique for all new oral drug candidates, where methods can be validated. This
strategic movement from plasma based analysis to DBS analysis has achieved a
significant step forward in the toxicological 3Rs (refined bleeding techniques,
reduced animal usage).
However, while organisations as a whole will realise these benefits, the switch
to DBS analysis would seem to provide little advantage for the bioanalyst. Non-optimised
automation, low sensitivity and increased ion suppression compared to plasma
assays are some of the disadvantages that can be encountered with DBS analysis.
The Bioanalytical Science and Development group at GSK is investigating direct
analysis techniques for DBS samples that could readily counter these
disadvantages, resulting in a process that is simpler than that currently used
for plasma analysis. It has been demonstrated that such techniques not only have
the potential to make significant time and cost savings in sample preparation,
but also to significantly increase assay sensitivity, potentially allowing DBS
analysis to include a wider range of compounds. Quantitative analysis using
direct analysis has also been demonstrated without the use of chromatographic
separation.
Evaluation of experimental drugs by
bioanalysis: reduction of needed animals
Ben Westerink and Thomas IFH
Cremers (Groningen Research Institute of Pharmacy, RUG and Brains-on-line BV, NL)
Development of experimental drugs aimed at the central nervous sytems requires
detailed information about their levels in brain and blood. In addition
information about the effects of the compounds on the biochemistry of the brain
is needed. For a full pharmacokinetic and pharmacodynamic evaluation of an
experimental drug, levels in blood as well as in the extracellular space are
required. In addition changes of the brain biochemistry are to be recorded. To
that end we use - in freely moving animals - chronic implantations of blood
cathethers and microdialysis probes or sensors in laboratory rats or mice.
To reduce the number of experimental animals for these experiments we have
followed 2 strategies:
a. The established method to sample extracellular space is microdialysis.
However the method lacks quantitative information as the recovery of the
compounds through the dialysis membrane is low and unpredictable. Methods to
circumvent this disadvantage require high amounts of experimental animals. We
have recently developed a probe that samples exogeneous and endogeneous
compounds with a recovery close to 100%.
b. A sensitive derivatization method was developed that enables simultaneous
determination by LC-MS of a large series of amino acid and amine-derived
neurotransmitters and related metabolites.
Combination of both methods reduces to amount of animals needed to about 10% of
the traditional approach.
Whole blood analysis with LC-MS/MS
using direct injection
Jaap Verweij (Schering-Plough,
Toxicology and Drug Disposition, Oss,
NL)
Mass
spectrometric developments in the last decade enable (sub)nanomolar detection of
drug compounds in biological matrices in a few microliters of blood. However,
the sampling and especially the handling of these small blood volumes is not
straightforward. We studied the feasibility of a recently developed 'sorbent
sampling technique' to handle these small blood volumes and the application to
support pharmacokinetic (PK) screening programs. This technique applies 5-10mL
of blood on a fibrous material packed into a cartridge. Blood samples absorbed
on these cartridges are eluted directly, on-line onto a solid-phase extraction
liquid chromatography/tandem mass spectrometry (SPE-LC/MS/MS) system. It is
shown that the sorbent sampling technique can be applied for a range of drug
compounds. In spite of issues with recovery and sample clean-up that need
further improvement, the sorbent sampling technique provided similar data as
compared to conventional analytics. The technique was successfully applied to
derive kinetic data from individual mice, thereby decreasing the number of
required mice for a PK study from 21 to 3.
Addressing the Challenges of Limited
Sample Volumes in Bioanalysis with Microscale LC-MS/MS
Paul Rainville (Waters, UK)
Small sample volumes from tail-bled rodents or pediatric studies place an extra
challenge on the bioanalytical scientist; how to achieve the desired levels of
sensitivity from limited sample volumes. The use of microscale separations have
shown potential for the high-sensitivity analysis of limited-volume samples in
the fields of proteomics and metabolite profiling. However, they have
traditionally required a very experienced analyst and specialized instrument
configurations. In this presentation, we will discuss the use of capillary-scale
(300 µm) UPLC coupled to tandem quadrupole MS to achieve pg/mL levels of
sensitivity from just a few microliters of sample. The analysis was performed on
a standard tandem quadrupole high-sensitivity mass spectrometer with an
easy-to-configure and exchange dedicated nano-spray source using a ceramic
tile-based separations with integrated emitter. The design of the source and
nano-tile correctly positions the separations device and emitter such that no
user intervention is required. In these experiments with sample derived from rat
plasma, the separations device showed excellent robustness towards the samples
with greater than 1,000 injections obtained and maintained the chromatographic
efficiency, such that the peak widths were similar or superior to standard
microscale LC/MS.
Radioactive
measurement: Combustion or Dissolving - Radioactive measurement of feces, blood,
liver, brain or fat tissue, a Comparison of Dissolving with Combustion
E. Beek-van Velzen (Solvay Pharmaceutical
Research Laboratories, C.J. van Houtenlaan 36, 1381 CP Weesp, The Netherlands)
Combustion of tissue is the most used technique to determine the amount of
radioactivity. This combustion technique is very labour intensive and therefore
a search for alternatives was made. Feces and blood were directly dissolved with
SolueneŇ 350. Liver, brain and fat were dissolved with SolueneŇ 350 and the
Adaptive Focused Acoustics wave technique of Covaris.
The solubility technique gave a recovery of 95-110% of the values found with
combustion.
Bioanalytical Method
Development for Therapeutic Peptides
Erin Chambers (Waters Corporation)
Developing bioanalytical methods for therapeutic peptides represents a
significant challenge for many laboratories, particularly those that have
historically worked with small molecules. This poster discusses the streamlining
and the transition from small molecule bioanalysis to peptide bioanalysis by
introducing logical step-wise MS, LC, and sample preparation method development
approaches. Important considerations for chromatography of peptides for
bioanalysis and an LC screening protocol are described. A highly selective SPE
screening method which provides significant sample concentration while
eliminating evaporation was developed based on 2 sorbents and 1 simple starting
protocol. Results from 12 diverse peptide therapeutics extracted from human
plasma are then provided as demonstration of the effectiveness of the
methodologies outlined in the application. Emphasis is placed on maximum
resource utilization and developing methods which are selective and robust
enough to meet regulatory criteria typically associated with small molecule
LC/MS/MS assays.
A bioanalytical method for the peptide desmopressin is developed using the
screening methods and run through partial validation as an example of successful
implementation of these concepts. The method uses 300 µL of human plasma and
easily achieves sensitivity in the pg/mL range.
A Leap Forward in
Sensitivity and Selectivity for PAH Analysis using GC-APLI-TOF/MS
Rob van der Heijden1, Patrick
van Houts1, George McLeod2, Thomas Arthen-Engeland3,
Carsten Baessmann3, Armin Holle3 (1. Bruker Daltonics BV, PO Box 88, 1530 AB, Wormer, The Netherlands,
2. Bruker Daltonics, Coventry, United Kingdom and
3. Bruker Daltonik GmbH, Bremen, Germany)
Most polycyclic
aromatic hydrocarbons (PAHs) containing more than three aromatic rings are
considered to be genotoxic and carcinogenic to humans. Therefore, PAHs are
classified as priority pollutants to be controlled in food and in the
environment. PAH analysis has not been performed effectively with atmospheric
pressure ionisation (API) mass spectrometry due to poor ionisation performance
with standard ESI and APCI.
We present a major step forward in analytical capability for PAH analysis though
Atmospheric Pressure Laser Ionisation (APLI). A novel APLI source is described
for coupling of chromatography to Bruker TOF or Qq-TOF mass spectrometers.
Efficient and highly selective ionisation is combined with a mass spectrometer
offering stable and precise measurements of mass and isotopic patterns, plus the
necessary dynamic range. The result is highly specific detection with
order-of-magnitude improvements in sensitivity over existing techniques.
In this paper we report on an analytical method for the determination of the
16 PAHs by GC/APLI-MS with the aim to transfer this method into routine
analysis. We report characteristic analytical method parameters and show
applications in a (spiked) river sediment sample and in a real-life fish liver
sample.
It is concluded that APLI-TOF-MS is a highly selective and efficient
technique for PAH analysis. The high selectivity of ionization for PAHs enables
ready detection in dirty samples. In addition, the formula-based identification
based on accurate mass and isotope pattern gives greatly enhanced identification
over GC-q-MS.
Improved Peptide
Identification with an Ultra High Resolution Quadrupole Time of Flight MS (maXis)
Patrick van Houts1, Rob van der
Heijden1, Markus Lubeck2, Marina Behrens2,
Andrea Schneider2, Carsten Baessmann2 and Oliver Raether2
(1. Bruker Daltonics, Wormer, The Netherlands and 2. Bruker Daltonik, Bremen, Germany)
Q-TOFs are
widely used working horses for proteomics due to their superior mass accuracy in
MS as well as in MSMS, making them ideal instruments for de novo sequencing and
PTM characterization. For peptide identification at very low concentrations or
for highly complex samples, MSMS sensitivity and duty cycle become especially
critical. A new designed ultra high resolution Q-TOF MS (maXis) shows
significant improvements for these types of samples.
In contrast to small molecule applications where MSMS sensitivity is usually
defined via signal to noise of a single fragment, peptide identification has
additional requirements: Completeness of a fragmentation pattern is more crucial
than optimized intensities for single masses. Thus generation of fragment ions
covering a broad mass range and their efficient transfer from the collision cell
into the TOF part of the instrument were optimized. Using a higher order
multipole as collision cell and an additional ion cooler for guiding ions into
the orthogonal acceleration showed a dramatic improvement for proteomics
applications.
As a model system for samples of high complexity, tryptic digests of E. coli,
separated using a 90 min nanoLC gradient were used, leading to 7000-10000
fragment spectra and more than 600 significant protein identifications from
100ng sample. For single protein digests, analyzed with nanoLC, high sequence
coverages could be obtained from sub fmol amounts.
Detection of
Metabolites in Urine at Microdosing Levels in Rats Using LC-MS/MS
Carmai Seto1, Tanya Gamble1,
Takeo Sakuma1, Jinsong Ni2, Fred Ouyang2, Devin
Welty2 and Andrew Acheampong2 (1. Applied Biosystems/MDS Analytical Technologies, 71 Four Valley Drive, Concord,
Ontario, Canada and
2. Dept of Drug Safety Evaluation, Allergan, Irvine, CA, USA)
Many new chemical entities fail in Phase I clinical trials due to poor
pharmacokinetics. Until Phase I, human pharmacokinetics can only be predicted
from animal models. Microdosing is an approach which may make human clinical
trails more effective, predictable and quicker because this approach allows for
an earlier assessment of the drugs pharmacokinetics and metabolism in human
sooner. Screening compounds for clinical trials using the microdosing approach
can become more effective with the combination of the QTRAP® 5500 LC/MS/MS
system and LightSight® software for metabolite identification.
Increased Selectivity
Without Loss of Sensitivity: Use of LC-MS3 for Sensitive Additional Structure
Information and LC-MRM3 for Highly Selective Quantification
Axel Besa and Jan Lembcke (Applied Biosystems Germany, Frankfurter Straße 129 B, 64293 Darmstadt, Germany)
Structure determination as part of metabolism and unknown ID necessitate
sensitive and selective product ion spectra. These easily can be obtained by
TripleQuadrupole instruments as well as IonTraps. Due to technical setup the
former suffers from bad duty cycle while IonTraps typically show weak low mass
fragments (70:30 low mass cut off). The latter can be bypassed by MSn
fragmentation at the expense of loss in sensitivity and increase in cycle time.
Nevertheless, these secondary or higher fragmentation pattern result in useful
further information for structural elucidation.
Beside the use of MSn experiments for gathering additional
structure information MS3 provides the capability to increase selectivity.
Commonly Multiple-Reaction-Monitoring (MRM) is dedicated sensitive and selective
mode for quantification on TripleQuad instruments. In complex matrices MRM
selectivity sometimes suffers from high background and hence the sensitivity of
method decrease. To prevent laborious development of new LC conditions the
increase of MRM selectivity would be beneficial and time-saving to meet desired
levels of sensitivity.
High Throughput
Screening of GSH Adducts Using Hybrid Linear Ion Trap Systems Coupled with Fast
Chromatography at Clinically Relevant Dose Concentrations
Hua-fen Liu1, Hesham Ghobarah2
, Renee Huang1, Jens Dahlmann3 and Elliott Jones1
(Applied Biosystems, part of Life Technologies, (1) Foster City, USA, (2)
Concord, Canada, (3) Darmstadt, Germany)
Trazodone is a triazolopyridine derivative which is
effective in the treatment of depressive disturbances, including depression
associated with anxiety and sleep disturbances. Therapy is usually started at a
total of 150 mg per day divided into two or three doses. Cmax was reached at 0.7
μg/ml (~1.82 μM), and 1.2 μg/ml (3.2 μM) after a single oral dose administration
of 75 mg and 150 mg prolonged release trazodone, respectively. Overdosage of
trazodone may cause an increased incidence or severity of any reported adverse
reactions. Trazodone is highly protein-bound and rare but severe hepatic
toxicity has been reported, which often described as idiosyncratic toxicity.
Trazodone is a 3A4 substrate and
extensively metabolized by hydroxylation in humans.
Several substrate glutathione adducts (GSH) were reported in human liver
microsomes at high substrate concentrations. Higher (10-50 μM) than clinical
relevant substrate concentration and complex sample preparation were often
needed in order to achieve the sensitivity desired. Investigation of GSH
formation at lower substrate concentration (1-3 μM) which is at clinical Cmax
relevant concentrations has not been explored due to the sensitivity limitation
of current technologies.
A number of QTRAP® specific methods are possible for GSH
detection. A combination constant neutral loss scan of 129 Daltons and a
negative precursor ion of 272 has also been suggested as a superior method for
GSH detection compared to single MS.
Historically, in order to detect all potential GSH adducts, two or more
injections were often used due to instrument scan speed limitations. The QTRAP®
5500 system is the first LC/MS with the capability to do both negative precursor
and positive neutral loss in one fast IDA (Information Dependent Acquisition)
scan cycle and Linear Ion Trap (LIT) scan speed up to 20,000 Da/sec with
superior sensitivity. Given these ground-breaking speeds, it is possible to
perform a combined PI/NL with polarity switching in IDA, with 4 dependent ion
trap MS/MS scans in approximately 2 seconds. Furthermore, the enhanced
resolution and enhanced that all GSH formations and potential structure product
ion scan ensure major information can be assigned from a single LCMS injection.
A second novel and highly effective method of GSH detection
possible on the QTRAP® is pMRM (predictive MRM). This method uses the next
generation MRM builder in LightSight software version 2.1 coupled with Excel™ to
create a complete set of potential MRM transitions for any GSH
biotransformation. In this case, a comprehensive set of greater than 80 phase I
transformations can be created in a GSH specific way to produce a positive MRM
list based on the neutral loss of 129 and a negative list based on the precursor
of 272. This pMRM GSH method is fundamentally a pos/neg switching IDA experiment
with 3 to 5 dependent LIT MS/MS scans. Since the MRM has a superior signal to
noise ratio compared to all other survey scan modes, this often returns the
greatest number of GSH confirmations in a single injection.
With these methods, a novel new approach with one injection as
a generic high-throughput screening method can be achieved with sufficient
sensitivity to ensure an accurate result at clinical relevant concentrations.
Simultaneous
quantification of testosterone and androstenedione in human serum using LC-MS/MS
E. Joos1,2, J. Van Bocxlaer2, M. Vandermarliere1
and K. Croes1 (1. Clinical Laboratory, AZ Groeninge, Reepkaai, B-8500 Kortrijk. Belgium
&
2. Laboratory for Medical Biochemistry and Clinical Analysis, Fac. Pharm. Sci.,
Ghent University, Harelbekestraat 72, B-9000 Gent. Belgium)
Testosterone (T) and androstenedione (AS) functionally belong to the androgen
class of hormones. Androgens have a variety of effects on reproductive as well
as non-reproductive tissues. T measurements consequently have many clinical
applications, in men (e.g. hypogonadism), as well as in children (e.g. pubertas
praecox) and women (e.g. hirsutism). T is extensively protein-bound in blood.
AS, at the other hand, is a weak androgen but a chief precursor for many other
androgens. AS measurements are equally diagnostic in many reproductive system
disorders but also in certain tumours. T, AS and other androgens are generally
quantified by immunoassay. Several literature reports, however, indicate that
immunochemical measurements of T, more particularly very low T concentrations in
woman, children and hypogonadal men, tend to overestimate the true
concentration. Also for AS, poor correlations were found between RIA and e.g.
LC-MS/MS. Cross-reactivity of closely related molecules and reduced overall
selectivity of immunoassays have initiated the search for, generally
chromatographic, alternatives.
Here, we describe the development of a fast and simple, fully validated LC-MS/MS
method for the simultaneous quantification of T and AS. 400 µL of serum is
extracted with diethylether and the extract is chromatographed using a Symmetry®
C8 (2.1x100mm, 3.5µm) column with a turnover time of 7 min. MRM detection of T
(m/z 289.2 > 96.9 and m/z 289.2 > 108.8), AS (m/z 287.2 > 96.9 and m/z 287.2 >
108.8) and their d3-labeled counterparts, used as internal standards, is
achieved on a Waters Quattro Premier triple quadrupole system. Calibration
curves (1/x weighting, in 0.1% BSA) were linear over the concentration range
0.05 to 15 ng/mL. Full validation according to the FDA guidelines revealed an
LOQ of 0.05 ng/mL for both androgens, an intra-run precision of ≤7.2% (T) and
≤9.3% (AS) as well as a between-day precision of ≤8.7% (T) and ≤9.9% (AS) at 4
different QC levels (in human serum, n=6), and (between-day, n=6) accuracy data
ranging from 97.0 to 105.8% (T) and from 100.5 to 105.3% (AS). External quality
control samples were also evaluated using our method (n=12 for T; n=10 for AS).
All but two results were situated within the (MS/MS method) mean ± 1 SD
interval, all results within mean ± 2 SD. No significant matrix effect was
demonstrated and recovery figures between 96.2 – 119.2% (T) and 99.7 – 122.8%
(AS) were found. Freeze-thaw, short-term and long-term stability were all
adequate. The impact of 4 different types of vacutainer tube was tested (n=5)
and found to be not significant (multiple test corrected one-way Anova).
Finally, the validated method was used in a method comparison with RIA (61 and
35 patient samples for T and AS, respectively). Passing-Bablok regression
revealed statistically different regression slopes, precluding the cross
platform use of reference intervals.
CHIRAL LC-MS-MS
DETERMINATION OF (R)- AND (S)-CITALOPRAM IN HUMAN PLASMA SAMPLES
E. Mallat, V. Jas, J. Flik, M. Boterman,
K. de Jonge (Analytisch Biochemisch Laboratorium (ABL) BV, Scholtenstraat 7, Assen, The
Netherlands)
Introduction
Drug monitoring of antidepressants is a required issue in order to achieve
optimal dosage of these compounds in patients and, therefore, avoid possible
intoxications or any other medical complications. Citalopram is a selective
serotonin reuptake inhibitor used as antidepressant. Its desired activity
resides mainly in the (S)- enantiomer and, in less degree, in its metabolite
(+)-(S)-desmethylcitalopram. In this work, the validation and application of a
sensitive and enantioselective LC-MS-MS method for the chiral separation and
quantitation of the enantiomers (R)- and (S)- citalopram in human plasma samples
is presented.
Material and Methods
After homogenization, centrifugation and addition of internal standard (d6-(RS)-citalopram)
to human plasma samples, liquid-liquid extraction at basic pH using t-butylmethylether
(TBME) was performed to isolate citalopram from the matrix. A chiral
chromatographic column CHIROBIOTIC V (Astec) modified covalently with
macrocyclic glycoproteins has been used to separate both enantiomers [1]. The
mobile phase consisted of a methanolic solution containing 0.05% formic acid and
0.01% ammonia. An API 4000 LC-MS-MS system (SCIEX) equipped with a Turbo Ion
Spray interface working in the positive mode was used as detector.
Results
The assay was validated over the concentration range from 1 to 500 ng/mL for
each enantiomer. Method validation was carried out to evaluate the accuracy and
precision of the calibration range, and the quality control samples, the
recovery of the enantiomers in plasma, the accuracy of the 10-fold dilution, and
the stability of the racemic mixture and of each of the enantiomers (stability
at 2-8oC, at room temperature, freeze/thaw stability and post-preparative
stability). Accuracy of the method (expressed as bias) of the quality control
samples was under 12%, whilst precision (expressed as C.V.) showed values from 2
to 5%. Stability studies revealed an acceptable low deviation (bias ≤15%) from
the target value for the racemic mixture as well as for each of the enantiomers.
Discussion and conclusion
A sensitive and enantioselective bioanalytical LC-MS-MS method for the
separation and quantitation of each enantiomer of citalopram has been developed
and validated successfully. This method has been applied satisfactorily to
determine (R)- and (S)- citalopram in clinical samples.
References
M. Kosel, C.B. Eap, M. Amey, P. Baumann (1998) J. Chromatogr. B 719 (1998)
234-238.
A sensitive LC-MS/MS
method for the separate quantification of gentamicin C1, C1a, C2 and C2a in
bovine tissue.
M. Boterman, M.J. Dröge, F. Oostebring, F. Venema
& E. Oosting (Analytisch Biochemisch Laboratorium (ABL) BV, Assen, The Netherlands)
Introduction
Gentamicin is a broad-spectrum aminoglycoside antibiotic, consisting of a
mixture of at least four components (C1, C1a, C2 and C2a) with three different
molecular weights. In veterinary medicine, gentamicin has been used since many
years for the treatment of bacterial infections in cattle. Therefore, tissues
and milk intended for human consumption have to be analysed for their maximal
residue limits (MRLs). The following MRLs have been established by the Committee
for Veterinary Medicinal Products (EMEA) for total gentamicin (sum of C1, C1a,
C2 and C2a) in bovine tissue: 200 µg/kg (liver), 750 µg/kg (kidney) and 50 µg/kg
(muscle and fat) [1]. Consequently, sensitive analytical methods are required to
determine these MRLs, and up till now, such a method was not available for the
individual determination of gentamicin C1, C1a, C2 and C2a in bovine tissue. The
ultimate goal of our work was to develop and validate a reliable and very
sensitive LC-MS/MS method suitable for the determination of gentamicin C1, C1a,
C2 and C2a in bovine tissue (kidney, liver, muscle and fat).
Materials and methods
After homogenisation of the tissue, the internal standard (tobramycin sulphate)
was added and a protein precipitation was performed using trichloroacetic acid.
After centrifugation, the samples were subjected to solid phase extraction (SPE)
using Isolute® CBA SPE columns (3 mL, 100 mg). The evaporated residues were
dissolved in injection solvent and injected into the LC-MS/MS system (Sciex
API4000). Gentamicin C1, C1a, C2 and C2a were fully separated using a Hypersil®
BDS C18 LC column (3 µm, 4.60 x 150 mm). The mass-spectrometer was equipped with
a Turbo Ion Spray interface and operated in the positive ion mode. A full
validation of the method was performed according the EC requirements [2]. The
calibration range covered 0.10 MRL – 10.0 MRL.
Results
The assay was validated in the concentration range of 75.0–7500 µg/kg (kidney),
20.0–2000 µg/kg (liver) and 5.00–5000 µg/kg (muscle and fat) total gentamicin.
The method showed acceptable accuracies (expressed as bias) and precisions
(expressed as CV) for all four components in all tissues and, typically, a lower
limit of quantification (LLOQ) of 0.10 MRL was achieved (Table 1). Moreover, the
method was found valid with respect to recovery, specificity and stability (48h
storage on the autosampler, one additional freeze/thaw cycle, 221 days long-term
storage at -70°C) for all matrices.
Discussion and conclusion
At ABL, a sensitive bioanalytical LC-MS/MS method for the separate
quantification of gentamicin C1, C1a, C2 and C2a in bovine kidney, liver, muscle
and fat has been developed and validated successfully. The method produced
accurate and precise results and very low LLOQs were achieved (down to 0.10 MRL)
in comparison to other available methods (usually ≥0.50 MRL). Due to the low
LLOQ and full separation of gentamicin C1, C1a, C2 and C2a, the assay has proved
to be extremely suitable for quantitative bioanalysis and has, therefore,
offered a competitive alternative to methods currently available for the
investigation of the depletion kinetics of gentamicin in bovine tissue.
Moreover, the method can be used as lead for the validation of the determination
of gentamicin in other species.
References
1. EMEA 2001. Committee for veterinary Products. Gentamicin. Summary Report.
2. Volume 8 of The Rules governing Medicinal Products in the European Union.
Notice to Applicants and Guideline. Establishment of maximum residue limits (MRLs)
for residues of veterinary medicinal products in foodstuffs of animal origin,
October 2005.
A sensitive automated online SPE-LC-MS/MS
method for the quantitative determination of ivermectin B1a in bovine plasma.
M. Boterman1, M.J. Dröge1,
F. Oostebring1, M.A.H. Boezeman2 and E. Oosting1
(1. Analytisch Biochemisch Laboratorium (ABL) BV, Assen, The Netherlands and 2. Farma Research Animal Health (FRAH) BV,
Nijmegen, The Netherlands)
Introduction
Ivermectin, a mixture of two homologous compounds ivermectin B1a and ivermectin
B1b, is a widespread used antiparasitic drug in veterinary medicine and is
routinely administered to millions of cattle per year. Over the years, the
topical “pour-on” formulation has largely displaced the conventional injectable
formulation in farming practices. Consequently, more sensitive analytical
methods are required for the quantitative determination of ivermectin for
pharmacokinetic purposes in bovine plasma samples. The objective of the present
study was to develop and validate a sensitive and reliable bioanalytical
LC-MS/MS method suitable for the determination of ivermectin B1a in bovine
plasma.
Materials and Methods
Bovine plasma samples were supplemented with an internal standard (abamectin)
and subjected to a protein precipitation using acetonitrile followed by online
solid phase extraction (SPE) on HySphereTM C8 EC-SE cartridges using a
SymbiosisTM Pharma system. Subsequently, the extracted samples were introduced
into the LC-MS/MS system for quantification. The samples were chromatographed on
a Zorbax Eclipse XDB-C8 column (3.5 µm, 75 x 4.6 mm). The mass-spectrometer
consisted of a Sciex API 4000 equipped with an atmospheric pressure chemical
ionization interface and was operating in the positive ion mode. A full
validation of the method was performed according to the current guidelines for
bioanalytical method validation [1,2]. The calibration range for ivermectin B1a
covered 0.200 – 200 ng/mL in bovine plasma. The validation included the
determination of the parameters: calibration, accuracy and precision, recovery,
specificity, dilution, and stability.
Results
The assay for ivermectin B1a was validated in the concentration range of 0.200 –
200 ng/mL and a Lower Limit Of Quantification (LLOQ) of 0.200 ng/mL was achieved
in bovine plasma. The method showed acceptable accuracies (expressed as bias)
and precisions (expressed as Intra- and Total CV) for all QC levels (Table 1),
which were determined in three validation runs. In addition, the method was
found valid with respect to recovery (protein precipitation step), specificity,
20-fold dilution of samples with blank matrix and stability (bench-top and
refrigerator stability (24 h.), on machine stability (72 h.) and stability after
three additional freeze/thaw cycles).
Discussion and Conclusion
At ABL, a sensitive bioanalytical LC-MS/MS method for the determination of
ivermectin B1a in bovine plasma has been developed and validated successfully.
In contrast to other available methods for the determination of ivermectin B1a
(usually with an LLOQ of 1.00 ng/mL), for the first time an automated online SPE
technique was used, resulting in highly accurate and precise results and a very
low LLOQ (0.200 ng/mL). Furthermore, the assay appeared to be extremely suitable
for the quantification of ivermectin B1a in bovine plasma samples and thereby
offered a competitive alternative to the methods available for the investigation
of the pharmacokinetics of ivermectin B1a in bovine plasma. In addition, it
should be noted that a similar method can be successfully applied for the
determination of ivermectin B1a in other species, such as horse, dog and cat.
References
1. Shah et al. Workshop/Conference report; Bioanalytical Method Validation. A
revisit with a decade of progress. Pharmaceutical Research 2000; 17: 551-1557.
2. Guidance for Industry, Bioanalytical Methods Validation. U.S. Department of
Health and Human Services, Food and Drug Administration. June 2001.
New UHPLC systems from Eksigent
Technologies for improved performance in LC/MS
Rein Reitsma (Separations Analytical Instruments)
Based on its Microfluidic Flow Control technology, Eksigent has recently
introduced a new Nanochip- and Micro LC system that can operate at pressures up
to 10,000 psi.
Our nanoLC-Ultra and cHiPLC systems have been developed to address the growing
demand for higher resolution separations, smaller samples and complex analyses
using nanoLC-MS. The increased maximum pressure allows for the use of longer
columns or columns packed with smaller diameter particles.
The ExpressHT-Ultra system is designed for fast LC/MS of small molecules using
microLC columns with an ID of 0.5 – 1 mm. Microfluidic Flow Control allows for
excellent control for gradients as short as 30 sec. A newly designed active wash
station ensures minimal carryover with cycle times as short as 1 min.
Performance equals or exceeds conventional UHPLC systems, while saving up to 90%
solvent. Applications of the new UHPLC system which demonstrate increased
throughput and performance with mass spectrometric detection are presented.
DRIED BLOOD SPOTS FOR
TK/PK AND BIOMARKER ANALYSIS AND 3R
Elwin Verheij (TNO Quality of Life, PO Box 360, 3700
AJ Zeist)
Dried blood spot sampling (DBS) as alternative to plasma collection contributes
to 3R in a relatively simple straightforward manner as result of improved animal
welfare (smaller sample volume) and reduction of the number of animals (less
satellite animals). Acceptance and widespread application of DBS depends on the
following issues (among others):
- quantitative data for TK/PK application
- very high sensitivity methods for PK of high potency drugs
- acceptance of whole blood values (plasma is the golden standard)
- compatibility with biomarker/omics analysis (disease/efficacy/safety)
- practical aspects, e.g. easy of use, robustness,..
We recently started investigating DBS for TK/PK, biomarker and
omics applications in life sciences research and in this paper we will present
some of our initial results.
Generic solid phase extraction
procedure for the quantitative analysis of peptides in plasma
H. Korthals, H. Gerritsen, R. Brust-van Schaik
and B. Buscher* (TNO Quality of
Life, Zeist, The Netherlands)
The quantitative analysis of peptides in plasma can be very challenging compared
to the bioanalysis of the more predictable ‘small molecules’. Method development
is often a matter of trial and error. Problems encountered include, amongst
others, adsorption of peptides to materials (glass, polypropylene,
polyethylene), instability of peptides in standard solution and/or in plasma and
strong binding of peptides to plasma proteins. A generic solid phase extraction
(SPE) procedure has been developed for the quantitative LC-MS/MS analysis of
peptides in plasma. The procedure is based on four different SPE materials and
procedures. All peptides investigated with the developed SPE procedure had a
recovery from plasma of > 80 % on at least one of the four SPE materials. Method
development including a short pre-validation takes approximately one week.
EX VIVO HUMAN SKIN MODEL FOR PHOTOGENOTOXICITY TESTING
AA Reus,
FAA van Acker, CAM Krul (TNO Quality of Life, Zeist, The Netherlands)
Photogenotoxicity testing aims at the early detection of the potential of
compounds to induce tumors upon activation with UV (photocarcinogenicity). In
contrast to standard genotoxicity testing, there are no equivalent in vivo
photogenotoxicity assays for additional evaluation of a positive (or equivocal)
result in vitro, or in vivo tests can not be used, as for personal care
products, due to EU legislation. Moreover, oversensitivity and the occurrence of
pseudo-effects with in vitro assays have become a major problem. Consequently,
the number of false positives and unnecessary in vivo photocarcinogenicity
studies will increase.
For these reasons, a relevant in vitro assay for evaluation of a photogenotoxic
potential in skin, was developed using ex vivo human skin.
Advantages of the human skin model are:
- best mimicking the human situation
- target organ for photocarcinogenicity
- relevant barrier properties of the stratum corneum
- metabolic capacity
Human skin is obtained from surgery. Circular membranes are cultured in an
air-liquid interface and exposed to test chemicals for 1 h, either via the
medium or the topical site of the skin, followed by UV exposure (5 J.cm-2 UVA,
0.19 J.cm-2 UVB) and preparation of cell suspensions for comet analysis.
Predictivity of the photo-comet assay was evaluated using 6 compounds (fluoroquinolones)
with varying degrees of photosafety liability. Sparfloxacin, lomefloxacin, and
ciprofloxacin (known photocarcinogenic potential) clearly demonstrated a
dose-related increase in %tail DNA in the presence of UV, while the non-photocarcinogenic
compounds levofloxacin, gemifloxacin and gatifloxacin did not. The %tail DNA was
not increased in the absence of UV. In parallel, the same compounds were
evaluated in the in vitro photo-comet with L5178Y cells and an in vivo skin
photo-micronucleus assay with rats, and compared with results of ex vivo human
skin to develop a tiered testing strategy for photogenotoxicity assessment. The
results with human and rat skin were comparable, while the in vitro photo-comet
appeared to be more sensitive.
Results thus far indicate that the human skin model can be a relevant
alternative for photogenotoxicity evaluation of compounds that reach the skin,
such as personal care products and pharmaceuticals.
Monitoring Matrix
EffectsM in Biological Samples Utilizing Dual Scan MRM Mode Mass Spectrometry
Paul D Rainville, Joanne Mather, Robert S Plumb,
Jing Lin (Waters Corporation, Milford, MA)
The process of developing LC/MS methodology for the purpose of quantitative
assays for medicines and their associated metabolites is a daunting challenge.
Human and animal biofluids such as urine, plasma and bile vary in components and
complexity. Additionally, endogenous analytes generated by a subject’s age,
gender and medication regimen can further interfere, contributing to the
complexity in quantitative bioassay development and validation. Chromatographic
techniques that utilize sub 2 micron particles have shown significant
contributions to the area of quantitative bioanalysis. This is due to the
increased chromatographic efficiency and subsequent generation of sharp
chromatographic peaks, often 2 – 3 seconds at base, resolving analytes of
interest from interfering matrix components.
However, even under the best chromatographic performance co-elution can occur.
Therefore implementing an approach whereby qualitative MS scan data obtained
from the matrix is simultaneously acquired with quantitative MRM MS data can aid
in the monitoring of potential interfering compounds ensuring assay robustness
and reproducibility. Furthermore, the fact that scan data is obtained
simultaneously with the MRM data means data mining can be performed at a later
time if questions arise as to the presence of analytes not targeted in the
original analysis.
Here we present the utilization of simultaneous acquisition of MS scan and MRM
data coupled with liquid chromatography utilizing sub 2 micron particle columns
as a method development tool for quantitative bioanalytical applications
The CMA 64 IView
microdialysis Catheter
Noud Grimberg (Aurora Borealis Control B.V.,
Postbus 2, 7760 AA Schoonebeek)
The CMA 64
IView microdialysis Catheter is a sterile, single use device intended for
intravenous monitoring through a standard peripheral vein catheter (PVC) up to
72 hours. The clean samples and possibility for frequent sampling makes it ideal
for research.
Small substances diffuse into the IView catheter reflecting accurate
concentrations in blood. It is perfused with a physiologic solution, containing
the anticoagulant dalteparin sodium, to avoid clotting
Accuracy of incurred
sample results, the final validation challenge in quantitative bioanalysis ?
Theo de Boer, Marjan Hagenouw, Rudy Francke and
Jaap Wieling (Xendo Drug Development B.V., Hanzeplein 1, NL-9713 GZ Groningen,
The Netherlands)
Since incurred
sample reproducibility validation has found a common ground in today’s
bioanalysis, a final challenge remains relatively unattended sofar, i.e.
accuracy of incurred sample results. One of the main reasons for this is the
disability to obtain or prepare a quality control sample that represents the
study samples, caused by the phenomenon that each individual sample from the
same and from different subjects is unique in its composition and hence its
behaviour in a bioanalytical method.
We propose a methodology that, similar to incurred sample reproducibility
determinations, utilises study samples to determine the accuracy of the results
obtained for them, by applying the standard addition approach and then determine
the spike recovery.
We have set up such an experiment for a bioequivalence study for alendronate.
Therefore, a relatively complex urine method involving derivatisation and
solid-phase extraction with LC-MS/MS quantitation with stable isotope labelled
alendronate as the internal standard was developed and validated prior to the
study sample analyses.
Subsequently study samples at three time points from several subjects were
analysed in singlicate. In two separate batches selected samples were repeated
in duplicate after additional spiking with analyte at two levels. These batches
were executed by two different technicians to include inter-analyst variation
The results enabled us to both evaluate incurred sample reproducibility but also
spike recovery and spike recovery repeatability. The spike recovery was then
considered to be a measure of incurred sample accuracy.
In general, the results for this approach looked very promising. For the samples
investigated by this approach generally the spike recovery was 80-120%, with
repeatability being less than 4% and reproducibility less than 15% for all
samples.
For this method applied to this study we showed that the spike recovery approach
to determine accuracy of incurred sample results performed satisfactory.
Thereby a proposed methodology is obtained to determine the accuracy of incurred
sample results, that may be used by the bioanalytical community to serve as a
basis for further discussions. We challenge others to build up experience with
this approach in order to be able to exchange experiences and further tune the
approach.
Application of Dried
Blood Spot sampling within clinical trials: Feasibility study of pardoprunox, a
potent anti-Parkinson Drug
O.A.M. Brockhoff1, P.H. van Amsterdam1,
J. Thomson2, C. Troupe3 (1. Solvay Pharmaceuticals, Global
Bioanalytics Department, Weesp, The Netherlands 2. Aptuit Inc, Riccarton,
Scotland, 3. Quintiles Laboratories, Marietta, GA, USA)
To assess
efficacy and safety of new drugs in development numerous trails in patients are
conducted. Often during these trails blood samples are collected to determine
drug and metabolite levels and/or biomarkers. Blood sampling, plasma or serum
generation, logistics and interim storage can be challenging and costly factors.
Dried blood sampling technique has been around for more then 40 years and is
widely used in newborn screening for the detection of inherited metabolic
diseases. As the technique could ease the sampling process in patient trails and
especially largely reduce the costs of sample logistics and storage there may be
great advantages to implement it. Obviously the analytes should still be
measurable with the same performance characteristics as in the ‘classical plasma
tube’ case.
To explore the possibilities of working with bloodspots, a feasibility study
has been done at Aptuit. The primary goal was to investigate blood spot analyses
of pardoprunox and its metabolites and to investigate stability at room
temperature. A key prerequisite was: limits of quantification must be comparable
to the bioanalytical method validated in human plasma.
|